IT Panel

Advantages of IT Panels (Isolated Power Panels):

Active Isolated Power Systems prevent the people in the environment from being harmed in case of any failure.
It prevents power cuts that may occur due to insulation failure.
The fault current in the electrical system is constantly monitored and reduced to non-critical levels.
It ensures the continuity of power supply for all medical areas, including the operating room, cardiac areas and intensive care units, which is the most critical of energy continuity.
Electrical fault monitoring is guaranteed.
With it panels, the risk of fire caused by leakage currents is eliminated.
Protection of high-cost medical devices against electrical failures is provided.
Additional leakage currents are collected and prevented from reaching critical values.
It is ensured that the operations continue without interruption.
Areas where IT Panels (Isolated Power Panels) are used in hospitals:
Operating theaters
Operating room preparation room, plaster room and recovery rooms
Intensive care units
Anesthesia rooms
Cardiac catheterization chambers
Angiographic examination rooms
Premature baby rooms
IT BOARDS (Medical Isolated Power Panels)

These panels are used in the power supply of places in medical locations within the scope of Group 2. Panels are comply with TS IEC 60364-7-710 standarts. Typical medical isolated power panels are consist of an isolating transformer, insulation monitoring device detect insulation resistance, load current, transformer temperature, system, earth, CT connections, alarm indicator, equipotential bonding terminal. Electrical loads in Group 2 locations are protected against short circuit currents with fuses appropriate to IT sistem.

it board

 Technical Specifications
 Brand  ELEKTROVIEN
 Type  MIPP / 1P-XX
 Standarts  TSE-IEC 60364-7-710
 Output Power  3,15 / 4 / 5 / 6,3 / 8 / 10 kVA
 Supply Input  Single phase Mono phase Line
 Rated Line Voltage  230 VAC
 Frequency  50 / 60 Hz
 Isolation Level  3 kV / 1 min.
 Input Protection  gL Fuse
 Output Voltage  230 VAC
 Output Protection  2 Pole Fuse
 Watchdog  IsolaIstion Resistance by LCD Screen
 Alarm output  Insulation Fault, Overload, Over temperature
 Functional Test  AdvancedInsulation Fault
 Enclosure Leakage Current  <0,5 mA
 Isolation fault detection period  <1 s
 Working Temperature  0°C / 50°C
 Storage Temperature  -15°C / 70°C
 Panel Dimensions  1700 x 500 x 500 mm
 Ventilation  with Fan
 Protection Class  IP 41
 Colour  RAL 9003 or RAL7035
 Distibution Output  6 - 12 - 18 - 24... Pcs


IT PANELS WITH TRANSFER UNIT

Medical isolated power panels with transfer unit; are used in the system is fed from two different power supply to provide power continuity of isolated power systems. In addition to isolated power panels, these panels have transfer unit provide switching between two paralel power source. Transfer unit follow two supply line and when voltage values exceed limit values, provide transfering between lines as automatically. In these way, interruption of energy is prevented and critical loads is always fed.

it board

 Technical Specifications
 Brand  ELEKTROVIEN
 Type  MITPP / 1P-XX
 Standarts  TSE-IEC 60364-7-710
 Output Power  3,15 / 4 / 5 / 6,3 / 8 / 10 kVA
 Supply Input  Double Single Phase Line
 Rated Line Voltage  230 VAC
 Frequency  50 / 60 Hz
 Isolation Level  3 kV / 1 min.
 Input Protection  gL Fuse
 Output Voltage  230 VAC
 Output Protection  2 Pole Fuse
 Watchdog  IsolaIstion Resistance by LCD Screen
 Alarm output  Insulation Fault, Overload, Over temperature
 Functional Test  Advanced Insulation Fault and Transfer Test
 Enclosure Leakage Current  <0,5 mA
 Isolation fault detection period  <1 s
 Working Temperature  0°C / 50°C
 Ventilation  with Fan
 Protection Class  IP 41
 Colour  RAL 9003 or RAL7035
 Transfer Time  < 5 ms.
 Response Range  50 - 500 k Ω
 Distibution Output  6 - 12 -18 - 24... Pcs
 Panel Dimensions  1700 x 500 x 500 mm


IT PANELS WITH TRANSFER UNIT AND FAULT DETECTION SYSTEM

Medical isolated power panels with transfer unit and insulation fault detection device have also test signal generator, insulation fault evaluators and toroidal transformer in apart from other isolated panels. When any fault exist, this fault is detected by insulation monitoring device and test signal generator produce a test signal, after that fault detected according to response of system to this signal. Insulation fault evaluator send signal to alarm monitoring devices.

it board

 Technical Specifications
 Brand  ELEKTROVIEN
 Type  MITFPP / 1P-XX
 Standarts  TSE-IEC 60364-7-710
 Output Power  3,15 / 4 / 5 / 6,3 / 8 / 10 kVA
 Supply Input  Double Single Phase Line
 Rated Line Voltage  230 VAC
 Frequency  50 / 60 Hz
 Isolation Level  3 kV / 1 min.
 Input Protection  gL Fuse
 Output Voltage  230 VAC
 Output Protection  2 Pole Fuse
 Watchdog  Isolation Resistance by LCD Screen
 Alarm output  Insulation Fault, Overload, Over temperature
 Functional Test  Advanced Insulation Fault and Transfer Test
 Enclosure Leakage Current  <0,5 mA
 Isolation fault detection period  <1 s
 Working Temperature  0°C / 50°C
 Storage Temperature  -15°C / 70°C
 Panel Dimensions  1700 x 500 x 500 mm
 Ventilation  with Fan
 Protection Class  IP 41
 Colour  RAL 9003 or RAL7035
 Transfer Time  < 5 ms.
 Response Range  50 - 500 k Ω
 Distibution Output  6 - 12 - 18 - 24... Pcs



ELEKTROVIEN